            "brand": "Datex-Ohmeda Inc",
            "score": 92,
            "title": "Giraffe Stand-alone Infant Resuscitation System; Powered emergency ventilator (resuscitator)\r\nProduct identifiers: 2082663-001, 2082663-002, M1241424\r\n\r\nProduct impacted only if service blender M1091607-R was installed during the last Preventative Maintenance",
            "errors": [],
            "hazard": "Blender knob shaft can loosen, which can affect the delivered oxygen concentration. The system can deliver an oxygen concentration that does not match the intended concentration set by the user, potentially resulting in hypoxia or hyperoxia in the patient.",
            "remedy": "Firm began notifying customers on 5/8/2026 via letters titled \"URGENT MEDICAL DEVICE CORRECTION.\"\n\nCustomers were instructed to identify affected devices by following instructions provided in the letter. Devices with affected blenders should be removed from patient use, and the customer should contact a GE HealthCare Service Representative.",
            "country": "US",
            "alert_id": "FDA-Z-2357-2026",
            "category": "Medical devices",
            "evidence": [
                {
                    "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfRES/res.cfm?id=220638",
                    "label": "FDA recall record"
                },
                {
                    "url": "https://api.fda.gov/device/recall.json",
                    "label": "openFDA device recall API"
                }
            ],
            "imageUrl": null,
            "recordId": "medical-device-alert-monitor:OPENFDA-DEVICE-RECALL:Z-2357-2026",
            "sourceId": "OPENFDA-API",
            "dedupeKey": "medical-device-alert-monitor:OPENFDA-DEVICE-RECALL:Z-2357-2026",
            "image_url": null,
            "matchedAt": "2026-06-23T06:46:58.094Z",
            "record_id": "medical-device-alert-monitor:OPENFDA-DEVICE-RECALL:Z-2357-2026",
            "riskLevel": "Open, Classified",
            "scrapedAt": "2026-06-23T06:46:58.094Z",
            "sourceUrl": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfRES/res.cfm?id=220638",
            "source_id": "OPENFDA-API",
            "confidence": 0.9,
            "dedupe_key": "medical-device-alert-monitor:OPENFDA-DEVICE-RECALL:Z-2357-2026",
            "matched_at": "2026-06-23T06:46:58.094Z",
            "risk_level": "Open, Classified",
            "runSummary": {
                "mode": "startUrls",
                "recordsSaved": 1,
                "sinceLastRun": false,
                "watchlistTerms": [],
                "sourceDiscovery": "openfda-device-recall-api"
            },
            "scraped_at": "2026-06-23T06:46:58.094Z",
            "sourceName": "OPENFDA-DEVICE-RECALL",
            "source_url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfRES/res.cfm?id=220638",
            "matchReason": "Matched openFDA device recall API record.",
            "productName": "Giraffe Stand-alone Infant Resuscitation System; Powered emergency ventilator (resuscitator)\r\nProduct identifiers: 2082663-001, 2082663-002, M1241424\r\n\r\nProduct impacted only if service blender M1091607-R was installed during the last Preventative Maintenance",
            "publishedAt": "2026-06-18T00:00:00.000Z",
            "retrievedAt": "2026-06-23T06:46:58.094Z",
            "run_summary": {
                "mode": "startUrls",
                "records_saved": 1,
                "since_last_run": false,
                "watchlist_terms": [],
                "source_discovery": "openfda-device-recall-api"
            },
            "source_name": "OPENFDA-DEVICE-RECALL",
            "match_reason": "Matched openFDA device recall API record.",
            "product_name": "Giraffe Stand-alone Infant Resuscitation System; Powered emergency ventilator (resuscitator)\r\nProduct identifiers: 2082663-001, 2082663-002, M1241424\r\n\r\nProduct impacted only if service blender M1091607-R was installed during the last Preventative Maintenance",
            "published_at": "2026-06-18T00:00:00.000Z",
            "retrieved_at": "2026-06-23T06:46:58.094Z",
            "scoreReasons": [
                "openFDA device recall API result",
                "Matched openFDA device recall API record.",
                "openFDA records are unvalidated and do not prove causation or replace medical advice."
            ],
            "score_reasons": [
                "openFDA device recall API result",
                "Matched openFDA device recall API record.",
                "openFDA records are unvalidated and do not prove causation or replace medical advice."
            ],
            "normalizedFields": [
                "title",
                "productName",
                "hazard",
                "remedy",
                "sourceUrl",
                "publishedAt"
            ],
            "normalized_fields": [
                "title",
                "product_name",
                "hazard",
                "remedy",
                "source_url",
                "published_at"
            ]
        },
        {
            "brand": "Datex-Ohmeda Inc",
            "score": 92,
            "title": "Giraffe Warmer with Integrated Resuscitation System (iRes), Panda Warmer with\nIntegrated Resuscitation System (iRes)\nProduct Identifiers: M1118179, M1112198\n\nImpacted only if service blender M1091607-R was installed during the last Preventative Maintenance.",
            "errors": [],
            "hazard": "Blender knob shaft can loosen, which can affect the delivered oxygen concentration. The system can deliver an oxygen concentration that does not match the intended concentration set by the user, potentially resulting in hypoxia or hyperoxia in the patient.",
            "remedy": "Firm began notifying customers on 5/8/2026 via letters titled \"URGENT MEDICAL DEVICE CORRECTION.\"\n\nCustomers were instructed to identify affected devices by following instructions provided in the letter. Devices with affected blenders should be removed from patient use, and the customer should contact a GE HealthCare Service Representative.",
            "country": "US",
            "alert_id": "FDA-Z-2358-2026",
            "category": "Medical devices",
            "evidence": [
                {
                    "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfRES/res.cfm?id=220639",
                    "label": "FDA recall record"
                },
                {
                    "url": "https://api.fda.gov/device/recall.json",
                    "label": "openFDA device recall API"
                }
            ],
            "imageUrl": null,
            "recordId": "medical-device-alert-monitor:OPENFDA-DEVICE-RECALL:Z-2358-2026",
            "sourceId": "OPENFDA-API",
            "dedupeKey": "medical-device-alert-monitor:OPENFDA-DEVICE-RECALL:Z-2358-2026",
            "image_url": null,
            "matchedAt": "2026-06-23T06:46:58.178Z",
            "record_id": "medical-device-alert-monitor:OPENFDA-DEVICE-RECALL:Z-2358-2026",
            "riskLevel": "Open, Classified",
            "scrapedAt": "2026-06-23T06:46:58.178Z",
            "sourceUrl": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfRES/res.cfm?id=220639",
            "source_id": "OPENFDA-API",
            "confidence": 0.9,
            "dedupe_key": "medical-device-alert-monitor:OPENFDA-DEVICE-RECALL:Z-2358-2026",
            "matched_at": "2026-06-23T06:46:58.178Z",
            "risk_level": "Open, Classified",
            "runSummary": {
                "mode": "startUrls",
                "recordsSaved": 2,
                "sinceLastRun": false,
                "watchlistTerms": [],
                "sourceDiscovery": "openfda-device-recall-api"
            },
            "scraped_at": "2026-06-23T06:46:58.178Z",
            "sourceName": "OPENFDA-DEVICE-RECALL",
            "source_url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfRES/res.cfm?id=220639",
            "matchReason": "Matched openFDA device recall API record.",
            "productName": "Giraffe Warmer with Integrated Resuscitation System (iRes), Panda Warmer with\nIntegrated Resuscitation System (iRes)\nProduct Identifiers: M1118179, M1112198\n\nImpacted only if service blender M1091607-R was installed during the last Preventative Maintenance.",
            "publishedAt": "2026-06-18T00:00:00.000Z",
            "retrievedAt": "2026-06-23T06:46:58.178Z",
            "run_summary": {
                "mode": "startUrls",
                "records_saved": 2,
                "since_last_run": false,
                "watchlist_terms": [],
                "source_discovery": "openfda-device-recall-api"
            },
            "source_name": "OPENFDA-DEVICE-RECALL",
            "match_reason": "Matched openFDA device recall API record.",
            "product_name": "Giraffe Warmer with Integrated Resuscitation System (iRes), Panda Warmer with\nIntegrated Resuscitation System (iRes)\nProduct Identifiers: M1118179, M1112198\n\nImpacted only if service blender M1091607-R was installed during the last Preventative Maintenance.",
            "published_at": "2026-06-18T00:00:00.000Z",
            "retrieved_at": "2026-06-23T06:46:58.178Z",
            "scoreReasons": [
                "openFDA device recall API result",
                "Matched openFDA device recall API record.",
                "openFDA records are unvalidated and do not prove causation or replace medical advice."
            ],
            "score_reasons": [
                "openFDA device recall API result",
                "Matched openFDA device recall API record.",
                "openFDA records are unvalidated and do not prove causation or replace medical advice."
            ],
            "normalizedFields": [
                "title",
                "productName",
                "hazard",
                "remedy",
                "sourceUrl",
                "publishedAt"
            ],
            "normalized_fields": [
                "title",
                "product_name",
                "hazard",
                "remedy",
                "source_url",
                "published_at"
            ]
        },
        {
            "brand": "Datex-Ohmeda Inc",
            "score": 92,
            "title": "Panda Freestanding Warmer, Infant Radiant Warmer\nProduct Identifiers: 2063568-001\n\nImpacted only if service blender M1091607-R was installed during the last Preventative Maintenance.",
            "errors": [],
            "hazard": "Blender knob shaft can loosen, which can affect the delivered oxygen concentration. The system can deliver an oxygen concentration that does not match the intended concentration set by the user, potentially resulting in hypoxia or hyperoxia in the patient.",
            "remedy": "Firm began notifying customers on 5/8/2026 via letters titled \"URGENT MEDICAL DEVICE CORRECTION.\"\n\nCustomers were instructed to identify affected devices by following instructions provided in the letter. Devices with affected blenders should be removed from patient use, and the customer should contact a GE HealthCare Service Representative.",
            "country": "US",
            "alert_id": "FDA-Z-2359-2026",
            "category": "Medical devices",
            "evidence": [
                {
                    "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfRES/res.cfm?id=220640",
                    "label": "FDA recall record"
                },
                {
                    "url": "https://api.fda.gov/device/recall.json",
                    "label": "openFDA device recall API"
                }
            ],
            "imageUrl": null,
            "recordId": "medical-device-alert-monitor:OPENFDA-DEVICE-RECALL:Z-2359-2026",
            "sourceId": "OPENFDA-API",
            "dedupeKey": "medical-device-alert-monitor:OPENFDA-DEVICE-RECALL:Z-2359-2026",
            "image_url": null,
            "matchedAt": "2026-06-23T06:46:58.219Z",
            "record_id": "medical-device-alert-monitor:OPENFDA-DEVICE-RECALL:Z-2359-2026",
            "riskLevel": "Open, Classified",
            "scrapedAt": "2026-06-23T06:46:58.219Z",
            "sourceUrl": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfRES/res.cfm?id=220640",
            "source_id": "OPENFDA-API",
            "confidence": 0.9,
            "dedupe_key": "medical-device-alert-monitor:OPENFDA-DEVICE-RECALL:Z-2359-2026",
            "matched_at": "2026-06-23T06:46:58.219Z",
            "risk_level": "Open, Classified",
            "runSummary": {
                "mode": "startUrls",
                "recordsSaved": 3,
                "sinceLastRun": false,
                "watchlistTerms": [],
                "sourceDiscovery": "openfda-device-recall-api"
            },
            "scraped_at": "2026-06-23T06:46:58.219Z",
            "sourceName": "OPENFDA-DEVICE-RECALL",
            "source_url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfRES/res.cfm?id=220640",
            "matchReason": "Matched openFDA device recall API record.",
            "productName": "Panda Freestanding Warmer, Infant Radiant Warmer\nProduct Identifiers: 2063568-001\n\nImpacted only if service blender M1091607-R was installed during the last Preventative Maintenance.",
            "publishedAt": "2026-06-18T00:00:00.000Z",
            "retrievedAt": "2026-06-23T06:46:58.219Z",
            "run_summary": {
                "mode": "startUrls",
                "records_saved": 3,
                "since_last_run": false,
                "watchlist_terms": [],
                "source_discovery": "openfda-device-recall-api"
            },
            "source_name": "OPENFDA-DEVICE-RECALL",
            "match_reason": "Matched openFDA device recall API record.",
            "product_name": "Panda Freestanding Warmer, Infant Radiant Warmer\nProduct Identifiers: 2063568-001\n\nImpacted only if service blender M1091607-R was installed during the last Preventative Maintenance.",
            "published_at": "2026-06-18T00:00:00.000Z",
            "retrieved_at": "2026-06-23T06:46:58.219Z",
            "scoreReasons": [
                "openFDA device recall API result",
                "Matched openFDA device recall API record.",
                "openFDA records are unvalidated and do not prove causation or replace medical advice."
            ],
            "score_reasons": [
                "openFDA device recall API result",
                "Matched openFDA device recall API record.",
                "openFDA records are unvalidated and do not prove causation or replace medical advice."
            ],
            "normalizedFields": [
                "title",
                "productName",
                "hazard",
                "remedy",
                "sourceUrl",
                "publishedAt"
            ],
            "normalized_fields": [
                "title",
                "product_name",
                "hazard",
                "remedy",
                "source_url",
                "published_at"
            ]
        }
    ]
}